Home > Water Quality > Model Q45D Dissolved Oxygen Monitor

Model Q45D Dissolved Oxygen Monitor
from Analytical Technology

Model Q45D Dissolved Oxygen Monitor

ATI's Model Q45D Dissolved Oxygen monitor is designed to provide reliable oxygen measurement and reduce maintenance costs. The galvanic dissolved oxygen sensor uses a rugged 5 mil Teflon membrane to resist mechanical abrasions and tears. The membrane serves as a barrier to allow molecular oxygen to diffuse into the reaction cell where it is reduced at the working electrode. This reaction produces a small current which is proportional to oxygen concentration.

In applications where process constituents adhere to the membrane, frequent membrane cleaning is required. To keep the dissolved oxygen system nearly maintenance free, an optional Auto-Cleaner can be included to remove those materials that build up on the sensor membrane. The cleaning cycle is user selectable to maximize the cleaning effectiveness.

When sensor maintenance is required, our cartridge-based sensor makes it fast and easy. Where other manufacturers make you throw away the cartridge, we've made our cartridge field repairable. We also include enough replacement parts to rebuild your cartridge ten times. When you consider the average cost of a competitor's cartridge at $125, ATI can save you thousands of dollars in maintenance costs.

Features

  • Loop-powered, AC or Battery Versions: This line of microprocessor based instrumentation allows for easy implementation of loop-powered or battery-powered capability within the same instrument. This instrument can be rapidly converted between any of these versions with no requirements for software change.
    • Loop-powered (16-35 VDC) Transmitter, 4-20 mA output
    • Line-powered (115/230 VAC) Analyzer, dual relays, dual 4-20 mA outputs
    • Battery-powered (( VDC) Monitor/Data Logger, dual 0-2.5 VDC outputs
  • Large, Dual Line Display: The large, high contrast, super-twist display provides excellent readability over a wide operating temperature range, even in low light conditions. The main display line consists of large, segmented characters with measurement units. The secondary display line utilizes easily readable dot matrix characters for clear display of calibration and diagnostic messages. Two or four measured parameters may be display simultaneously.
  • Interactive User Interface: Four button programming provides intuitive navigation through the menu driven user interface.
  • Dual Alarm/Analog Outputs: AC operated systems provide two relays that are configurable for either "control mode" or "alarm mode" or operation.
  • Flexible Calibration: Air and sample calibration options include temperature, barometric pressure, and salinity compensation. All calibration methods include stability monitors to check temperature and main parameter stability before accepting data.

Clean Running Means Better Performance

It's a fact: A clean D.O. sensor performs better

That's because when there is nothing between the sensing membrane and the water in the tank, you get uniform oxygen transfer for accurate measurement of dissolved oxygen content. And that translates into improved process performance, energy savings and more effective water treatment.

Trouble is, not all D.O. monitors are easy to clean. Units that do not clean themselves need to be removed for frequent maintenance. And those that do clean themselves with grindstones or brushes can be expensive and difficult to maintain. Still others use chlorine gas to kill contaminants on the sensor's surface, but do nothing to remove contaminant buildup, leaving the source of the problem intact.

Enter ATI's Auto-Clean D.O. monitor, the first D.O. monitor to clean itself with air. Biological growth and other contaminates are literally blasted from the Teflon membrane. The result: reduced maintenance, better performance and more accurate D.O. monitoring.

Keeping Clean Was Never So Simple

All dissolved oxygen sensors are affected by process coating. Any material that slows the diffusion of oxygen into the reaction cell will cause an error in the oxygen reading. If the system does not have a means to clean itself, the sensor will require frequent manual cleanings, which means higher maintenance costs.

Our self contained Auto-Clean system generates a series of high pressure air blasts to remove foulants from the face of the sensor. Each cleaning cycle lasts approximately 3 minutes, during which the monitor outputs are placed on HOLD condition to prevent false readings or alarms. The user can vary the cleaning cycle frequency from as often as once every two hours, to as little as once a day.

ATI was the first to introduce a dissolved oxygen system that cleans itself with air. Others have tried to copy our systems, but none are more effective at removing biological growth and other contaminants than the ATI Auto-Clean system.

Our Sensors Makes Sense

The Auto-Clean sensor uses the same rebuildable cartridge as our submersible style sensor. The durable 5 mil Teflon membrane allows the sensor to perform in the most demanding applications. Modular design lets you easily remove the sensing element for service, reducing maintenance costs. The sensor's principle of operation is simple. Oxygen diffuses through a Teflon membrane and reduces on the surface of a working electrode, generating a small electrical current proportional to D.O. concentration. An RTD temperature element measure water temperature and correct the sensor signal for its effect. The result is a D.O. measurement that is accurate over an operating range of 0-50°C.

The Heart of the System

The Auto-Clean monitor's electronic controls make operation easy, accurate and reliable:

  • LCD display of D.O. concentration and temperature
  • Dual isolated analog outputs that are assignable to D.O. and/or temperature
  • Microprocessor based electronics, with operator interface through four button keypad on the front panel
  • A security code to protect calibration and control settings from unauthorized tampering
  • Contact outputs that include two programmable control relays with variable deadband and time delay functions
  • One relay is used to control the Auto-Cleaner, the other can be configured for either control or alarm mode
  • Internal compressor provides high pressure air for cleaning system

System Installation

The installation of an Auto-Clean D.O. system is quick and simple. A special mounting adapter connects the sensor to standard 1" conduit or pipe and another adapter provides secure connection of the pipe to a standard handrail system. The NEMA 4x monitor and cleaner package may be wall mounted or connected to a standard handrail using vertical supports supplied as part of a handrail mounting kit. If desired, the D.O. monitor may be located up to 1,000 feet away from the cleaning system, which is always located within 50 feet of the sensor.

Rebuildable Cartridge Design

The sensor is constructed with a Teflon membrane that is used to isolate the electrochemical cell from the process. Oxygen that is dissolved in the process diffuses through the membrane and is reduced on the surface of the working electrode. The standard sensor uses a membrane with a 5 mil thickness to resist abrasion and tears in high solids applications. For processes that run at lower oxygen concentrations or require a faster response time, a 2 mil membrane is optional.

The electrochemical components of the sensor are contained in an easily removable cartridge. This modular design makes sensor maintenance fast and easy as the cartridge is easily regenerated by replacing the membrane and electrolyte. This saves you money because you do not have to throw the cartridge away each time you regenerate the sensor. To further reduce your costs and frustration, every sensor includes 10 replacement membranes and electrolyte, so now you do not have to wait for replacement cartridges

Multiple Power Sources

The Q45D is designed for exceptional flexibility to meet a variety of monitoring applications. In its simplest form, the unit is a loop-powered transmitter operating from standard 24 VDC power supplies. This means easy integration into many DCS, PLC and SCADA systems. Loop-powered instruments include the PID output functions.

For applications where a second analog output is desired or where alarm relay functions are needed, an AC powered system is available. Operation from AC power allows the user to utilize analog outputs for independent PID and temperature or simply dissolved oxygen and temperature. The AC powered system also includes 2 electrochemical relays that can be configured for control, alarm or timer modes.

For even greater versatility, a portable unit operating from a standard 9V battery is also available. This instrument can be supplied with an internal data logger, making it ideal for short term monitoring at remote sites. The unit will run for 10 days on a single battery and the data logger will store up to 32,000 data points, easily enough for 10 days of data at 1 minute intervals.

Specifications

Display Large 4 digit main display, 0.75" characters; 12 digit alpha-numeric second line display
Measurement Range 0 - 40,000 ppm
Accuracy 0.2% of span or better
Repeatability +/- 0.05% of span or better
Linearity 0.1% of span or better
Temperature Drift 0.01% of span/°C
Power 16-35 VDC for loop-powered unit 115/230 VAC, 50/60 Hz, 10 VA max; 9 volt battery for battery operating portable
Relay Mode Programmable for control, alarm, or timer function
Analog Output Isolated 4-20 mA, 550 ohms max load, Two assignable 4-20 mA outputs, 550 ohms max (AC only)
Control Relays Two SPDT relays, 6A @ 250 VAC, 5A @ 24 VDC, resistive
Data Logger Battery version only, stores 32,000 data points
Operating Temperature -20 to +60°C
Operating Humidity 0-95% non-condensing
Enclosure NEMA 4X (IP-66) polycarbonate wall, panel or pipe mount
Weight 15 pounds
Sensor Membrane covered galvanic sensor
Sensor Wetted Materials Noryl and 316 Stainless Steel
Sensor Cable 30 feet standard, 1,000 feet maximum
Sensor Response Time 90% in 60 seconds (2 mil membrane); 90% in 180 seconds (5 mil membrane)
Sensor Temperature Limits 0-50°C

Ordering Information

For pricing please call AFC International at 800-952-3293, 219-987-6825 or fax 219-987-6826 or email sales@afcintl.com.
New Products | Air Sampling | Gas Detection | Other Meters | Respiratory | Water Quality | WMD HazMat
Applications | Product Specials | Rentals | Links | Index | Contact Us | Home


AFC International, Inc.
PO Box 894, DeMotte, IN 46310
715 SW Almond St, Ste C, DeMotte, IN 46310
800-952-3293 219-987-6825 Fax 219-987-6826
eMail: sales@afcintl.com

Last Revised: 1/23/07