AFC International - Specialists in gas detection equipment, WMD / Hazmat equipment, Air sampling instrumentation, respiratory protection equipment and isocyanate detectors for product brands like Draeger and MultiRae Systems.  
 
 


Click here to view larger image


Mail E-mail this product to a friend


previous  up  next 

Model Q45P/R pH/ORP Monitor from Analytical Technology

ATI's versatile Q45P/R pH/ORP system is designed for use in all industrial and municipal applications. The sensors are engineered to function normally in applications where conventional sensors quickly fail.

Conventional pH/ORP sensors have an "open reference system", which means the reference element and electrolyte are in contact with the process. This allows chemicals to diffuse into the reference chamber and alter the reference system. As the reference junction becomes contaminated, the reference potential shifts. As chemicals attack the silver reference wire, the reference potential shifts and the sensor no longer functions. When any of these problems occur, cleaning and calibration cycles increase and the sensors fail much faster, all of which costs you time and money.

The differential pH/ORP sensors utilize a "sealed reference system", where the sensors are constructed with a second glass pH electrode as the reference element. The glass reference system protects the sensor from chemical poisons (sulfide, cyanide, chlorine, bisulfates, etc.) that destroy conventional pH sensors.

A large volume, dual junction saltbridge is used to maximize the in service time of the sensor. The annular junction provides a large surface area to minimize the chance of fouling. Large electrolyte volume and dual junctions minimize contamination of the reference solution. The replaceable saltbridge allows for easy and inexpensive sensor regeneration.

As integral preamplifier is encapsulated in the body of the sensor. This creates a low impedance signal output which ensures stable readings in harsh environments and maximizes the distance between sensor and analyzer. Sensor diagnostics are used to alarm the user in the event of electrode breakage, loss of sensor seal integrity or integral temperature sensor failure.

Sensor electrodes can be user specified to ensure measurement reliability and maximum sensor lifetime. The type of glass used in the pH electrodes can be selected for optimal performance. The metal electrode used for ORP measurement can be platinum or gold, depending on the chemical makeup of the process solution.

All sensors are available in several mounting configurations. The 1" NPT convertible style sensor is constructed of PEEK, a high performance thermoplastic that provides outstanding mechanical strength and chemical resistance. An insertion style sensor is available in 316SS for hot tap applications. The mounting hardware for the insertion sensor is either CPVC of 316SS and is mounted using a 1 ¼" NPT full ported ball valve. The sensors are also available with 1 ½" or 2" sanitary style fittings.

Auto-Clean Option - The Q45P/R Auto-Clean System is designed to extend sensor cleaning intervals on pH and ORP sensors that are mounted in applications containing high levels of solids contamination. The Auto-Clean System uses high pressure air to remove contaminants from the face of the sensor automatically.

The Auto-Clean system utilizes a self contained compressor and air cylinder to generate a series of high pressure air blasts. Each cleaning cycle lasts approximately 2 minutes, during which the monitor outputs are placed in a HOLD condition to prevent false readings or alarms.

The concentration of solids in the process will determine the cleaning frequency of the system. The user can vary the cleaning cycle frequency from as often as once very two hours to as little as once a day.

Although the Q45 Auto-Clean System can greatly extend sensor cleaning intervals, the user must still clean and calibrate the sensor at some interval. For a typical pH or ORP measurement in wastewater treatment, the system can extend weekly cleaning intervals to once every 3 months. Cleaning systems are not suitable for removing hard scale such as that caused by lime.

Standard Sensor Compatibility - The Q45P/R monitors can be easily configured to accept the input from a traditional combination sensor. For applications where a specialty sensor is required, ATI offers a line of standard sensors to meet your needs.

The unique "Lock-N-Load" system is pipe mounted and provides quick and easy isolation of the sensor from a flowing stream. The self sealing sensor holder allows the sensor to be removed for maintenance without shutting down the process stream.

For high purity water (conductivity below 20 microSiemens), a special sensor is used in conjunction with a stainless steel flow assembly. This system is designed to minimize interference (noise, streaming potentials, CO2, etc.) and provides a stable pH measurement.

Sensor Features

Differential Style Sensors: The reference is a second glass pH electrode immersed in a reference buffer solution. This glass reference system allows the sensor to be used in applications that poison conventional pH sensors

PEEK Sensor Body Construction: All structural components are made of PEEK, a high performance thermoplastic that provides outstanding mechanical strength and chemical resistance.

Replaceable Sensor Saltbridge: A large volume junction saltbridge is used to maximize the in service time of the sensor. The replaceable saltbridge allows for easy and inexpensive sensor regeneration.

Sensor Diagnostics: Sensor diagnostics are used to alarm the user in the event of electrode breakage, loss of sensor seal integrity or integral temperature sensor failure.

Multiple Sealing Materials: Multiple sealing materials are used to preserve sensor integrity over a wide range of chemical processes and varying temperatures

Integral Preamplifier: The integral preamplifier is encapsulated in the body of the sensor. This creates a low impedance signal output which ensures stable readings in harsh environments and maximizes the distance between sensor and analyzer.

Monitor Features

Loop-powered, AC or Battery Versions: This line of microprocessor based instrumentation allows for easy implementation of loop-powered or battery-powered capability within the same instrument. This instrument can be rapidly converted between any of these versions with no requirements for software change.

  • Loop-powered (16-35 VDC) Transmitter, 4-20 mA output
  • Line-powered (115/230 VAC) Analyzer, dual relays, dual 4-20 mA outputs
  • Battery-powered (( VDC) Monitor/Data Logger, dual 0-2.5 VDC outputs

Sensor Compatibility: Q45 monitors can accept standard pH sensor inputs, allowing easy upgrade for existing installations.

Diagnostic Messaging: Diagnostic messages provide a clear description of system conditions, which eliminates confusing error codes that are difficult to decipher.

Flexible Calibration: Two point and sample calibration options include auto-buffer recognition for 13 built in buffer tables. Manual overrides of the automatic buffer values for user to customize values. All calibration methods include stability monitors to check temperature and main parameter stability before accepting data.

Dual Alarm/Analog Outputs: AC operated systems provide two relays that are configurable for either control, alarm or operation.

Standard PID Output: A PID control output is standard in every Q45 pH or ORP monitor, with control parameters easily user configurable.

Monitor Specifications

Display Parameters Main Input: pH - 0 to 14
ORP - 1000 to +2000 mV
Sensor Voltage: pH - ±500 mV
Loop Current: 4.00 to 20.00 mA
Sensor Temperature: -10 to 110°C
Main Parameter Range 0.00 to 14.00 pH, -1000 to +2000 mV
Input Impedance >1013 Ohms
Repeatability 0.1% of span or better
Sensitivity 0.05% of span
Non-Linearity 0.1% of span
Stability 0.05% of span per 24 hours
Temperature Drift Span or zero, 0.02% of span/°C
Warm Up Time 7 seconds to rated performance
Response Time 6 seconds to 90% of step input at lowest setting
Max Sensor to Analyzer Distance 3000 feet w/preamp, 25 feet w/o preamp
Enclosure NEMA 4X, IP66, polycarbonate weatherproof and corrosion resistant, 4.9" x 4.9" x 5.5"
Mounting Options Wall, panel, pipe/handrail
Conduit Openings Three M16 threaded entries, 3 cord grips and 2 conduit adapters included
Weight 2 wire or battery: 1 pound
AC power units: 2 pounds
Auto-Clean units: 15 pounds
Display Large, high contrast LCD 4 digit main display with signs, 0.75" seven segment characters, 12 digit alpha numeric second line display, backlight
Ambient Temperature Service: -20 to 60°C
Storage: -30 to 70°C
Ambient Humidity 0 to 95% non condensing
Location Designed for non hazardous areas
EMI/RFI Influence Designed to EN 61326-1
Output Isolation 600V galvanic isolation
Filter Adjustable, 0-9.9 minutes additional damping to 90% step input
Temperature Input Selectable Pt1000 or Pt100 RTD with automatic compensation
Control Relays Two SPDT relays, 6A @ 250 VAC, 5A @ 24 VDC resistive
Analog Output Isolated 4-20 mA, 550 ohm maximum load. Two assignable 4-20 mA outputs, 550 ohm max
Relay Mode Programmable for control, alarm or timer function
Power 16-35 VDC for loop powered unit; 115/230 VAC, 50/60 Hz 10 VA max 2 "AA" batteries for battery operated portable
Data Logger Battery version only, stores 32,000 data points

Sensor Specifications

Measuring Range 0 to 14 pH, -1000 to + 2000 mV
Sensitivity 0.002 pH; 0.2 mV
Stability 0.02 pH; 2 mV (per 24 hours, non cumulative)
Wetted Materials PEEK, ceramic, titanium, glass, Viton, EDPM
Platinum or gold (ORP only)
316 SS with sanitary or insertion body styles
Temperature Compensation Pt1000 RTD
Sensor Cable 6 conductor plus 2 shields, HDPE jacket
Temperature Range -5° to 95°C
Pressure Limits 0-100 psig
Max Flow Rate 10 feet per second
Max Sensor to Analyzer Distance 3,000 feet
Sensor Body Options 1" NPT convertible, 1 ¼" insertion, 1 ½" or 2" sanitary style

Ordering Information

For pricing please call AFC International at 800-952-3293, 219-987-6825 or fax 219-987-6826 or email sales@afcintl.com.

Description
Catalog No.
Model Q45 pH Monitor, 24 VDC
Q45-P-1-1
Model Q45 pH Monitor, 115 VAC with 2 Relay & Outputs
Q45-P-2-1
Model Q45 pH Monitor, Auto-Clean 115 VAC & Heater
Q45-P-6-1
Model Q45 ORP Monitor, 24 VDC
Q45-R-1-1
Model Q45 ORP Monitor, 115 VAC with 2 Relays & Outputs
Q45-R-2-1
Model Q45 ORP Monitor, Auto-Clean, 115 VAC & Heater
Q45-R-6-2

Downloads

Model Q45P/R pH/ORP datasheet




 

Related Products


Model Q22 Specialty pH/ORP Sensors from Analytical Technology
Model Q22 Specialty pH/ORP Sensors from Analytical Technology
Series A15/76 Turbidity Monitor from Analytical Technology
Series A15/76 Turbidity Monitor from Analytical Technology
Series A15/66 Residual Sulfite Monitor
Series A15/66 Residual Sulfite Monitor
Q45H/64 Dissolved Ozone Monitor by ATI
Q45H/64 Dissolved Ozone Monitor by ATI

Other Recommended Products


Model PQ45 Portable Monitor System from Analytical Technology
Model PQ45 Portable Monitor System from Analytical Technology
 
Model Q45H/62-63 Residual Chlorine Monitor from Analytical Technology
Model Q45H/62-63 Residual Chlorine Monitor from Analytical Technology
 
Model Q45H/79 Total Chlorine Monitor from Analytical Technology
Model Q45H/79 Total Chlorine Monitor from Analytical Technology
 
Model Q45WQ Water Quality Monitor from Analytical Technology
Model Q45WQ Water Quality Monitor from Analytical Technology
 
 
 

Home | About Us | Our Products | Applications | Repair | Resources | Contact Us | Site Map | Login
Products Categories: New Products | Product Specials | Gas Detection | WMD/Hazmat | Isocynate | Air Sampling | Respiratory | Water Quality | Other Instruments | Rental
Product Industries: Industrial Hygiene | Environmental | Confined Space Entry | Fire/Hazmat | Indoor Air Quality (IAQ) | Water Quality | WMD / Law Enforcement

   Popular Product Searches: Ammonia Gas Detector |  Benzene Gas Detector |  Chlorine Gas Detector |  Combustible Gas Detector |  VOC Gas Detector
   Popular Brand Searches: Analytical Technology |  Bacharach Inc |  Crowcon Detection Instruments |  Draeger Safety |  Environics |  GfG Instrumentation |  Honeywell Analytics | 
Rae Systems |  Sensit Technologies |  Sensor Electronics |  Viasensor

AFC International, Inc.
Tel: 800-952-3293 or 219-987-6825 Fax: 219-987-6826
sales@afcintl.com
715 SW Almond St, Ste C | DeMotte, IN 46310
PO Box 894 | DeMotte, IN 46310

Website Maintained by Evolution Point, LLC.

 

Website maintained by Evolution Point - chicago marketing agency, web development chicago, chicago web development, print design, integrated marketing services chicago, interactive agency, search engine optimization, multilanguage website

Web Analytics