Nitrous Oxide Monitor from Bacharach Inc.

The Bacharach TGA (Trace Gas Analyzer) is an advanced N2O Analyzer. It uses a highly accurate non-dispersive IR cell providing superior stability and repeatability, with a measurement resolution of 1 ppm - full range. It provides measurements in either ppm, real time or elapsed time TWA and it has a leak check mode with audible and visual alarms. It has an internal sampling pump and the self contained pruge-air bag provides a constant zero enabling the analyzer to be used in contaminated environments. The TGA has data logging capabilities to store up to 200 test which can be downloaded to a printer or to a computer.

The rechargeable Ni-cad battery provides 8 to 12 hours of continuous operation with a low battery indicator. It provides an effective means of monitoring occupational exposure to N2O in hospital operating rooms, dental offices, veterinary clinics and surgical care centers. The TGA can also be used to check for leaks in medical gas pipelines.

Specifications

Range 0-1,000 ppm N2O
Resolution 1 ppm full range
Warm up Time 15 minutes
Accuracy +/- 1 ppm up to 100 ppm reading
+/- 5% of reading over 100 ppm reading
Case Metal with fabric pouch
Power Source Ni-cad rechargeable battery with gas gauge
Operating Temperature 32-122°F
Relative Humidity 0-99% non condensing
Dimensions 8"L x 19"W x 5"H
Weight 6 pounds, 12.4 ounces
Warranty Instrument: 1 year
IR Cell: 2 years (excluding cell contamination)

Ordering Information

Each unit includes a charger. For pricing, please call AFC International at 1-800-952-3293, 219-987-6825 or fax 219-987-6826 or email sales@afcintl.com
Description
Catalog No.
TGA Monitor for N2O
3015-4790
Inlet Filter
3015-3420
Printer
24-1229

Downloads

Features

Continuous data logging, stores up to 200 readings

Two TWA capabilities - real time & elapsed time

Built in purge cycle using a bladder bag provides a constant zero

Displays ppm/TWA simultaneously

Internal pump for remote sampling with low flow indicator

Built in, user programmable alarm

Rapid detection of changing N2O levels

Easy to read LCD display

8-12 hours of operation on full charge

Maintenance free infrared N2O sensor