The A24 Puffer is a portable, battery operated instrument containing a gas generator and simple hand pump that provides the puff of air needed for sensor response checks.
For puff testing diffusion gas detectors, the Puffer includes a short outlet tube. Gas is ejected from the generator by pressing the bladder on the front of the unit. For testing detectors with internal sampling pumps, a plug on the unit is removed so that the sampling pump can pull air through the generator cell for delivery to the detector.
Puffers are water resistant for use in any environment and are designed to intrinsic safety standards. Generator cells provide hundreds of tests, with an LED to indicate the remaining life of the generator cell. Generator cells are easily changed and the life indicator can be reset when a new generator is installed.
Puffer units are available for testing the following gas sensors:
Specifications
| Power |
9 VDC Alkaline battery |
| Generator |
Electrochemical |
| Generator Life |
1000 Cycles for Chlorine, 350 cycles for other gases |
| Pump |
Manual bladder |
| Display |
Single digit LED |
| Outlet Tube |
6" polyethylene |
| Enclosure |
Nema 4X, IP-66 |
| Size |
4.7" x 3.2" x 2.4" |
| Weight |
0.7 pounds |
Ordering Information
For pricing, please call AFC International at 1-800-952-3293, 219-987-6825 or fax 219-987-6826 or email
sales@afcintl.com. Each Puffer bump test generator is supplied with a specific gas generator cell factory installed. Spare generator cells may be ordered separately. Spare generators should be stored in a refrigerator in the sealed container in which they are shipped.
| Description |
Catalog No.
|
| Puffer for Chlorine |
00-1397
|
| Chlorine Generator Cell |
00-1433
|
| Puffer for Hydrogen Sulfide |
00-1398
|
| Hydrogen Sulfide Generator Cell |
00-1436
|
| Puffer for Ammonia |
00-1399
|
| Ammonia Generator Cell |
00-1434
|