Home > Gas Detection Instruments > Chip Measurement System

Chip Measurement System
from Draeger Safety

Chip Measurement System

The System

The CMS combines substance specific measuring Chips with an electronic based Analyzer for an easy to use spot-measurement system. The 10 capillaries in each Chip are filled with a reagent system based on Draeger's 60+ years of colorimetric chemistry expertise. The analyzer combines a revolutionary optical system for analyzing the color reaction with a mass flow controller and pump system that provides unbeatable reproducibility.

Fast Response Time

The CMS Analyzer requires no warm-up time. After taking a test, the results can be reported as quickly as 20 seconds. Higher concentrations, which pose a greater hazard, are reported faster than trace background levels. This increases the safety of using the CMS system.TWA and action levels are typically analyzed within 1-2 minutes. After providing the measurement results, the CMS is immediately ready for another test or another Chip for further analysis of the environment.

The expertise of Draeger's worldwide experience in industrial gas measurement has come to focus in the CMS. This is combined with a photo-optical controller that measures the speed of the chemical reaction, thus eliminating the human objectivity. Unlike other systems that rely on the volume of air, the CMS pulls the same mass of air through the capillary every time. The mass flow controller automatically compensates for changes in atmospheric pressure.

Accuracies of ±4 to 7% measured value is achieved for most gases. CMS has been validated by a nationally accredited third party laboratory

CMS Chip

Chemical Specific Chips

When you need to know exactly what is there, CMS is the tool to provide you those answers. The chemistry is chosen to provide the best results possible. The capillaries of several chips have multiple levels (Yes, even in something that small!) to help eliminate cross sensitivity or derive a specific measurement of the targeted chemical.

The CMS Chip

The Chip consists of 10 measurement capillaries filled with substance-specific reagent systems. The gas type, part number and batch number are printed on the chip. A printed barcode on the chip, read by the analyzer optics, contains information about gas type, measuring range and measuring time required for completing the measurement. Each chip is calibrated during manufacturing and is valid for two years.

Chip Measurement System

Simplified Operation

The CMS is literally as easy to operate as 1-2-3. Making analysis is as simple as turning the unit "on" (moving the slide position to position 1), then following the plain language instructions on the display to load the Chip and move the slide selector to positions 2 and 3. Operating the CMS could not be easier because this is the exact same procedure for every Chip! Upon completion of analysis, the CMS Analyzer shows the concentration in the display. Any misinterpretation of the results is eliminated by the unambiguous digital display of the concentration.

Simplified Maintenance

The CMS Analyzer does not require gas calibration! All calibration information is stored on the bar code of the CMS Chip and automatically downloaded to the analyzer when inserted. An electronic system self test is performed every time the Analyzer is switched ON, and a sampling system test is performed before each test. You are always assured that the CMS system is operating properly. Operating on alkaline batteries means that there are no worries about charging or battery memory. CMS is gas detection made simple.

Data Recorder

The internal data recorder stores up to fifty measurements with the time/date, contaminant/concentration and the sequential number of the measurement in a series. Measurement location information (in alphabetical code) may be entered manually. The stored data can be recalled and displayed on demand.

CMS Emergency Response Kit

Accessories

In order to measure inaccessible areas, the analyzer can be equipped with a remote system which consists of an additional pump and extension hose. The remote system can be installed in the back of the analyzer. Once attached, the remote system is activated by a separate switch and can remain mounted to the analyzer indefinitely. Additionally, a telescopic wand may be used for remote sampling.

A kit for emergency response is now available. It includes an analyzer, remote system, telescopic probe, 10 chips, flashlight/lantern, notepad and a chip case. Note: Chips marked with a (*) are included with this kit.

Specifications

Size 8" x 3.6" x 1.7"
Weight 25.6 ounces
Ingress Protection IP-54, protected against dust and water
Data Capacity Stores up to 50 measurements
Power Supply 4 x 1.5 Volt (AA) alkaline batteries
Battery Duration 450 minutes of analysis time (typically greater than 100 measurements)
Approvals UL: Class I, Div 1, Groups A, B, C, D, T4
CSA: Class I, Div 1, Groups A, B, C, D, T4
CENELEC: EEx ib IIC T4
MSHA, permissible gas analyzer
Operational Temperature 32-104°F

Ordering Information

Each monitor includes calibration adapter, tubing, manual and case. Call for information on the 450 Series with vibrating option. For pricing, please call AFC International at 1-800-952-3293, 219-987-6825 or fax 219-987-6826 or email sales@afcintl.com.
Description
Catalog No.
CMS Analyzer Set, includes analyzer, allen wrench and batteries
6405300
CMS Analyzer Set with Remote System
8317700
Remote System for measurement of inaccessible place, includes 3M of hose with adapter, screwdriver & float probe
6405060
CMS Emergency Response Kit includes analyzer, remote system, telescopic probe, 10 chips (*) and carrying case
4055711
CMS Emergency Response Kit without Chips, includes analyzer, remote systems, telescopic probe and carrying case
4055976
CMS Indoor Air Quality Kit, includes analyzer, batteries, nylon trasnport case and 1 each of Carbon Dioxide, Carbon Monoxide, Formaldehyde and Ozone Chips
4056455
Telescopic Probe with Viton Sampling Hose, remote system required
8316530
CMS Sampling Hose with Adapter and Float Probe, remote system required
8317614
CMS Extension Hose Adapter
4056713
Training Chip
6406290
Soft Side Case for analyzer and chips
4594631
Molded Plastic Transport Case
4056442
CMS Emergency Response Case with foam inserts, holds 14-16 chips
4055979
CMS Handbook
9092086
CMS Training CD
4056637

Chips

Chip
Measuring Range
Catalog No.
Acetic Acid
2.0-50.0 ppm
6406330
Acetone
40.0-600 ppm
6406470
Ammonia
0.20-5.0 ppm
6406550
Ammonia
2.0-50.0 ppm
6406130
Ammonia*
10.0-150 ppm
6406020
Ammonia
100-2,000 ppm
6406570
Benzene
0.05-2.5 ppm
6406600
Benzene(**)*
0.20-10.0 ppm
6406030
Benzene(**)
0.50-10.0 ppm
6406160
Benzene
10.0-250 ppm
6406280
Butadiene
1.0-25.0 ppm
6406460
Carbon Dioxide
200-3,000 ppm
6406190
Carbon Dioxide*
1,000-25,000 ppm
6406070
Carbon Dioxide
1.0-20% vol
6406210
Carbon Monoxide*
5.0-150 ppm
6406080
Chlorine*
0.20-10.0 ppm
6406010
Ethanol
100-2,500 ppm
6406370
Ethylene Oxide
0.40-5.0 ppm
6406580
Formaldehyde
0.20-5.0 ppm
6406540
Hydrochloric Acid*
1.0-25.0 ppm
6406090
Hydrochloric Acid
20.0-500 ppm
6406140
Hydrogen Cyanide
2.0-50.0 ppm
6406100
Hydrogen Peroxide
0.20-2.0 ppm
6406440
Hydrogen Sulfide
0.2-5.0 ppm
6406520
Hydrogen Sulfide*
2.0-50.0 ppm
6406050
Hydrogen Sulfide
20.0-500 ppm
6406150
Hydrogen Sulfide
100-2,500 ppm
6406220
Iso-Propanol
40.0-1,000 ppm
6406390
Mercaptan
0.25-6.0 ppm
6406360
Methanol
20.0-500 ppm
6406380
MTBE
10.0-200 ppm
6406530
Methylene Chloride
20.0-400 ppm
6406510
Nitrogen Dioxide*
0.50-25.0 ppm
6406120
Nitrous Fumes*
0.50-15.0 ppm
6406060
Nitrous Fumes
10.0-200 ppm
6406240
Oxygen
1.0-30.0% Vol
6406490
Ozone
25.0-1,000 ppb
6406430
Perchloroethylene*
5.0-150 ppm
6406040
Petroleum Hydrocarbons
20.0-500 ppm
6406200
Petroleum Hydrocarbons
100-3,000 ppm
6406270
Phosgene
0.05-2.0 ppm
6406340
Phosphine
0.10-2.5 ppm
6406400
Phosphine
1.0-25.0 ppm
6406410
Phosphine
20.0-500 ppm
6406420
Phosphine
200-5,000 ppm
6406500
Propane
100-2,000 ppm
6406310
Styrene
2.0-40.0 ppm
6406560
Sulfur Dioxide
0.40-10.0 ppm
6406110
Sulfur Dioxide
5.0-150 ppm
6406180
Toluene
10.0-300 ppm
6406250
Training Chip
-
6406290
Trichloroethylene
5.0-100 ppm
6406320
Vinyl Chloride
0.30-10.0 ppm
6406170
Vinyl Chloride
10.0-250 ppm
6406230
Water Vapor
0.40-10.0 mg/L
6406450
Xylene
10.0-300 ppm
6406260

(**)Please Note: The two Benzene chips in the CMS program have nearly the same measurement range. The primary difference is the humidity limits. The Benzene 0.5-10 ppm chip is designed for high humidity environments. Please refer to the following table.

Maximum Humidity,
Benzene 0.2-10 ppm
Temperature °F
Maximum Humidity,
Benzene 0.5-10 ppm
100%
32
100%
100%
50
100%
100%
68
100%
99%
86
100%
76%
95
100%
60%
104
80%

Performance Data

Measuring Range & Resolution Depends on the chip type
Typical Measuring Time 1-2 minutes in the range of the limit values, dependent on the chip type and the concentration of the hazardous substance
Contamination Effects Not possible
Calibration Not necessary
Operational Temperature 34 to 104°F (0 to 40°C)
Storage Temperature; Analyzer -4 to 140°F (-20 to 60°C)
Storage Temperature; Chips <77°F (<25°C)
Atmospheric Pressure 700 to 1,100 HpA (or Mbar)
Humidity 0 to 95% relative, not condensed
Determination of the Measurement Value Optics & light transfer system reflective measurement
System Diagnostics Self-performing, microprocessor controlled for all components of the system
Display LCD, 2 x 16 alphanumeric digits with backlight
Menu Languages English, German, French, Spanish
Operational Capacity Approximately 450 minutes on one set of batteries
Power Source 4 x 1.5V (AA) alkaline batteries
Weight 1.6 pounds (730g)
Dimensions 8" x 3.6" x 1.7"
Approvals UL Class 1, Div. 1, Groups A, B, C, D, Temp Code T4
Protection Class Dust & water spray protection according to IP54
New Products | Air Sampling | Gas Detection | Other Meters | Respiratory | Water Quality | WMD HazMat
Applications | Product Specials | Rentals | Links | Index | Contact Us | Home


AFC International, Inc.
PO Box 894, DeMotte, IN 46310
715 SW Almond St, Ste C, DeMotte, IN 46310
800-952-3293 219-987-6825 Fax 219-987-6826
eMail: sales@afcintl.com